stored. The locations of the buttons are shown with red rectangle frames
in Figure 4-29.
Figure 4-29
Remote Interface Operation
You can use the following commands from your PC terminals to store or
retrieve readings in the memory
In addition, the number for STORE
.
function only can be set through front panel.
INITiate
T his command tells the meter to be on "wait-for-trigger" state.
(
After a measurement is taken, measurement readings will be placed in
the memory.
)
FETCh?
U se this command to retrieve stored readings.
(
)
DATA:POINts?
U se this command to query the number of stored
(
readings
. )

4.4.4 Sensitivity Band (Hold)

The reading hold function captures and holds a stable reading on the
display. The multimeter beeps and holds the value when it detects a
stable reading. The sensitivity band in reading hold decides which
reading is stable enough. This band is expressed as a percent of reading
on the selected range. The multimeter considers a reading stable when
three consecutive values are within the band. You have an option to
adjust this sensitivity band.
Default
99 99 99 99