PERiod? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONTinuity?
DIODe?
TCOuple
?
TEMPerature?
The CONFigure Command
The CONFigure command offers a little more flexibility than the MEASure?
Command. The multimeter sets the parameters for the requested function,
range and resolution, but does not make the measurements. You have an
option to change the configuration. To initiate the measurement, use INITiate
or READ? Command.
CONFigure:
VOLTage:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
VOLTage:DC:RATio {<range>|MIN|MAX|DEF },{<resolution>|MIN|MAX|DEF}
VOLTage:AC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CURRent:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CURRent:AC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
RESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
FRESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
FREQuency {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONTinuity
DIODe
TCOuple
TEMPerature
CONFigure?
The READ? Command
The READ? Command changes the state of the trigger system from the
"idle" state to the "wait-for-trigger" state. When the specified trigger
condition requirements are met after the multimeter receives the READ?
command, the measurement will be initiated. The results are sent to the
120
120
120
120