Remote Interface Operation
Use the following commands to enable and configure MX+B function:
CALCulate:FUNCtion MXB
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:MXB:MMFactor {<value>|MINimum|MAXimum}
CALCulate:MXB:MMFactor? [MINimum|MAXimum]
CALCulate:MXB:MBFactor {<value>|MINimum|MAXimum}
CALCulate:MXB:MBFactor? [MINimum|MAXimum]
4.3.7 dB/dBm
A. dB
Definition
The dB feature takes a DC or AC voltage measurement and displays it in
decibel unit in correspondence to a relative reference value. The
calculation of dB is listed below:
V
=
×
dB
20
log(
in
)
Vref
or
dB = (Input signal in dBm) – (relative value in dBm)
NOTE: The
is the input signal and the
is the relative reference.
※
V
V
in
ref
The dB measurement is applied to DC and AC voltage only. The relative
value is adjustable and is stored in a volatile memory. The range for the
relative value is between 0 dBm and
200
dBm.
How to set the relative value
You can either manually select a number for the relative reference or let
the multimeter take the first reading (0 dB).
Front Panel Operation
Press CONFIG+ SHIFT+ NULL buttons, and then use
and
to move
◁
▷
through the digits and
and
buttons to increase or decrease the
△
▽
numbers to your desired value. Press ENTER to set the value. The
90 90 90 90