Picotest M3500A User Manual page 168

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

register, all conditions will be reported in the standard event summary
bit. At the moment, you have to write a decimal value by using the *ESE
command to set the enable register mask.
Note 1: The errors in the DMM's error queue except reading the error
queue by using SYSTem:ERRor? will be recorded by an error condition
(including bits 2, 3, 4 or 5 of the standard event register).
Note 2: In both the standard event register bit 3 and the questionable
data event register bits 0, 1 or 9, a reading overload condition can be
reported all the time. But in the DMM's error queue, no error message
will be recorded.
Bit Definitions for Standard Event Register
Bit
0.
Operation Complete
1.
Not Used
2.
Query Error
3.
Device Error
4.
Execution Error
5.
Command Error
6.
Not Used
7.
Power On
The following conditions will clear the standard event register.
You send a *CLS command.
You query the event register by using the *ESR? command.
The following conditions will clear the standard event enable register.
You turn on the power, and have set the DMM previously by using
the *PSC 1 command.
Decimal
Value
All commands prior to and including an *OPC command have been
1
executed
Set to 0.
2
The DMM output buffer was empty as the DMM tried to read from it.
4
Or a new command line was received before a previous query has
been read. Or, both the input and output buffers are full.
8
Error occurrence from a self-test, calibration or reading overlaod.
16
Error occurrence from an execution.
32
Error occurrence from a command's syntax.
64
Set to 0.
Since the last time the event register was read or cleared, power has
128
been turned off and on.
Definition
168
168
168
168