Status Reporting Commands - Picotest M3500A User Manual

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

The following conditions will clear the questionable data event register.
You execute a *CLS command.
You query the event register by using STATus:QUEStionable:EVENt?.
The following conditions will clear the questionable data enable register.
You turn on the power without using *PSC command.
You execute the STATus:PRESet command.
You execute the STATus:QUEStionable:ENABle 0 command.
C.10 Status Reporting Commands
SYSTem:ERRor?
Query the multimeter's error queue. Up to 20 errors can be stored in the
queue. Errors are retrieved in first-in-first out (FIFO) order. Each error
string may contain up to 80 characters.
STATus:QUEStionable:ENABle <enable value>
Enable bits in the Questionable Data enable register. The selected bits are then
reported to the Status Byte.
STATus:QUEStionable:ENABle?
Query the Questionable Data enable register. The multimeter returns a
binary-weighted decimal representing the bits set in the enable register.
STATus:QUEStionable:EVENt?
Query the Questionable Data event register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
STATus:PRESet
Clear all bits in the Questionable Data enable register.
*CLS
Clear the Status Byte summary register and all event registers.
170
170
170
170

Advertisement

Table of Contents
loading

Table of Contents