Choosing USER takes you to a menu where you can specify factors used
in the calculation equation to obtain the temperature. Use
move through the digits and
desired value. Press ENTER to set the value.
Choosing SPRTD takes you to a menu where you can specify the seven
coefficients that are used to determine the temperature. Use
to move through the digits and
desired value. Press ENTER to set the value.
Remote Interface Operation
Use the following commands to set up the RTD configuration.
SENSe:TEMPerature:RTD:TYPE{PT100|D100|F100|PT385|PT3916|USER|SPRTD|NTCT}
SENSe:UNIT {Cel|Far|K}
SENSe:UNIT?
SENSe:TEMPerature:RTD:RZERo {<value>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:ALPHa {<value>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:BETA {<value>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:DELTa {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:RZERo {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:A4 {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:B4 {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:AX {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:BX {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:CX {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:DX {<value>|MINimum|MAXimum}
SENSe:TEMPerature:TRANsducer FRTD
SENSe:TEMPerature:TRANsducer RTD
Thermocouple
Definition
If you are using the thermocouple function, the selections on the M3500A
are: type E, J, K, N, R, S and T. (On the other hand, other selection
"SIMULATED" simulates the reference junction.) To use the the
and
to change the numbers to a
and
to change the numbers to a
67 67 67 67
and
to
and