3.5 Continuity Measurements

M3500A uses 1 K Ω range for the continuity measurement. The meter
beeps when the test resistance is less than the threshold resistance. The
default threshold resistance is 10 , but you can set the threshold
resistance to anything between 1
by you is stored in a volatile memory and will be cleared after the meter
has been turned off. The source current for the continuity measurement
is 1 mA.
Warning! The maximum input voltage allowed is 1000 V. Applying
excess voltage may damage the meter and cause unpredictable
hazards.
How to measure the continuity
Selects input signal connections on front or rear panel.
Connect the test leads to the terminals as shown in Figure 3-2.
① ① ① ①
Set REConnect the input signal as the following figure.
② ② ② ②
Set threshold resistance via pressing CONFIG + CONT buttons. (Or
③ ③ ③ ③
skip this step if default resistance is to be used) When ready, press
ENTER button.
Press CONT button.
④ ④ ④ ④
The measured value will be shown on the display automatically. Meter
⑤ ⑤ ⑤ ⑤
"beeps" when measured resistance value is lower than threshold
value.
5
4
3
Figure 3-10
and 1 K
. The resistance value set
2
1