Picotest M3500A User Manual page 66

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

Where:
If you are using SPRTD(Standard Platinum RTD), select SPRTD and
specify the seven coefficients under SPRTD submenu.
The ITS (International Temperature Scale) -90 standard provides two
reference equations for Standard Platinum Thermometers covering
temperature range from 18.8033K to 1234.93K. However, one SPRTD
usually cannot cover the entire range, so the temperature range is
broken up into a few subranges. These subranges depend on the
calibration point of temperature scale and are based on the melting or
triple points of various pure substances. For an exact list of the elements
needed and details on RTD calibration, refer to NIST Technical Note 1265
"Guidelines For Realizing the International Temperature Scale of 1990".
In each subrange, the calibration constants required for that subrange
are listed.
Default
The default sensor type, units and transducer in M3500A is PT100,
and 4W RTD in order.
How to set up RTD
You can set up the RTD configuration either through the front panel
operation or through the remote interface operation.
Front Panel Operation
If you are using RTD, press CONFIG and then TEMP. Use
locate SENSOR submenu. Press ENTER to go to the submenu. Use
and
to locate your sensor type Press ENTER to select your desired
sensor type. So are the Units and Transducer settings.
(
2
=
+
+
1
R
R
At
Bt
t
0
δ
α
=
+
A
1
100
4
αδ
=
B
10
8
αβ
=
10
C
66 66 66 66
)
and
to