3.7.1 Thermocouple Measurements
Connect the thermocouple adapter to the banana jacks on the front
panel. The difference between each type is subject to the patch thermal
leads.
Note: Only connection via the front panel can be used for temperature
※
measurements.
How to measure thermocouple
Use terminals switch to select terminals on the front panel.
① ① ① ①
To measure thermocouple, you must use a thermocouple adapter and
② ② ② ②
a sensor, (K type for example), with the low thermal patch lead as
shown in Figure 3-12.
Configure the thermocouple type and unit using CONFIG + SHIFT +
③ ③ ③ ③
TEMP, such as K TYPE and ºC. When ready, press ENTER button.
Press SHIFT + TEMP buttons.
④ ④ ④ ④
Take the reading on the display.
⑤ ⑤ ⑤ ⑤
3.7.2 RTD Measurements
There are three kinds of temperature measurements with RTDs: 2-wire,
3-wire and 4-wire measurements. You will find connection instructions
and measuring procedures in the following sections.
5
4
3
Figure 3-12
44 44 44 44
1
2