2.
Press
button to set the number of readings. Use
▽
buttons to move between the digits and
increase or decrease numbers as desired. When ready, press ENTER
button.
Note: The MEM annunciator will be lit while the multimeter stores
※
readings and be turned off when the specified number of readings is
reached.
How to recall the stored readings
Press SHIFT + STORE buttons, and the multimeter will display all the
stored readings starting from the first reading. Use
move from the first reading to the last reading the meter has stored.
Remote Interface Operation
You can use the following commands from their PC terminals to store or
retrieve readings in the memory
function only can be set through front panel.
INITiate
T his command tells the meter to be on "wait-for-trigger" state.
(
After a measurement is taken, measurement readings will be placed in
the memory.
)
FETCh?
U se this command to retrieve stored readings.
(
DATA:POINts?
readings
. )
4.4.4 Sensitivity Band (Hold)
The reading hold function captures and holds a stable reading on the
display. M3510A/11A beeps and holds the value when it detects a stable
reading. The sensitivity band in reading hold decides which reading is
stable enough. This band is expressed as a percent of reading on the
selected range. M3510A/11A considers a reading stable when three
consecutive values are within the band. You have an option to adjust this
sensitivity band.
.
U se this command to query the number of stored
(
a nd
▽
◁
In addition, the number for STORE
79 79 79 79
and
◁
▷
buttons to
△
and
buttons to
▷
)