Picotest M3500A User Manual page 77

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

Defaults
The default of the trigger delay is automatic. M3500A automatically
selects a delay time for you according to the setting of the measurement
if you do not specify a delay. A list of the default for each measurement
function is shown on Table 4-4. The range for the delay is from 0 to
3600 seconds. User's selection of delay time is stored in a volatile
memory and the default will be restored when the meter is power-off.
The internal settings of automatic trigger delay
The delay time for automatic trigger delay is determined by the selection
of measurement function, range, and the setting of integration time and
AC filter speed. The following table shows the default delay time
for each measurement setting.
Measurement Function
DCV/DCI
Ω2 and Ω4
(PLC >= 1)
Ω2 and Ω4
(PLC < 1)
ACV/ACI
(Remote Interface/
External Trigger/
Single Trigger)
ACV/ACI
Table 4-4
Setting
PLC >= 1
PLC < 1
100Ω ~ 100kΩ
1 MΩ
10 MΩ ~ 100 MΩ
100Ω ~ 100kΩ
1 MΩ
10 MΩ ~ 100 MΩ
3 Hz
20 Hz
200 Hz
3 Hz
77 77 77 77
Trigger Delay
Time
1.5 ms
1.0 ms
1.5 ms
15 ms
100 ms
1.0 ms
10 ms
100 ms
7.0 s
1.0 s
600 ms
1.5 s