Limits Test - Picotest M3500A User Manual

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

CALCulate:NULL:OFFSet
< value>|MAXimum|MINimum
{ { { {
} } } }

4.3.5 Limits Test

The limits testing operation allows you to adjust a maximal and a
minimal limit values. The multimeter beeps and an "HI" or "LO" message
will be shown when the reading exceeds the upper or lower limit
respectively. You can specify the limit values and the values are stored in
a volatile memory. The default values for both upper and lower limits are
"0". This function is available to all except continuity and diode
measurements.
N ote: You can also use this HI/LO measurement as a Pass/Fail
application through the USB interface's Pin 2 and Pin3. But the USB
interface has to be turned off by pressing MENU
> ▷ > I N T E R F A C E >
.
E N T E R > ▷ > I N T E R F A C E > E N T E R > U S B > E N T E R > D I S A B L E > E N T E R
How to set the limits
You can set the limits or make a limit testing either through the front
panel or the remote interface operation.
Front Panel Operation
To set the limits, you need to configure the LIMITS math function by
pressing CONFIG + SHIFT + RATIO, and then use
and
to locate
either lower limit (MIN) or upper limit (MAX). Press ENTER to select the
limit you wish to set. Then use
and
to move through the digits and
and
to increase or decrease the numbers to a desired value. Press
ENTER to set the value. The locations of the buttons are shown with red
rectangle frames in Figure 4-18.
85 85 85 85

Advertisement

Table of Contents
loading