resolution. The reading is sent to the output buffer.
MEASure:VOLTage:DC:RATio? {<range>|MIN|MAX|DEF },{<resolution>|MIN|MAX|DEF}
Preset and make a dc:dc ratio measurement with the specified range and
resolution. The reading is sent to the output buffer. For ratio measurements, the
specified range applies to input signal, yet autorange is selected for the
reference signal.
MEASure:VOLTage:AC? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make an AC voltage measurement with the specified range and
resolution. The reading is sent to the output buffer. For AC measurement,
resolution is fixed at 6 ½ digits. Therefore the resolution parameter only affects
the front panel display.
MEASure:CURRent:DC? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a dc current measurement with the specified range and
resolution. The reading is sent to the output buffer.
MEASure:CURRent:AC? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a AC current measurement with the specified range and
resolution. The reading is sent to the output buffer. For AC measurement,
resolution is fixed at 6 ½ digits. Therefore the resolution parameter only affects
the front panel display.
MEASure:RESistance? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a 2-wire Ω measurement with the specified range and
resolution. The reading is sent to the output buffer.
MEASure:FRESistance? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a 4-wire Ω measurement with the specified range and
resolution. The reading is sent to the output buffer.
MEASure:FREQuency? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a frequency measurement with the specified range and
resolution. The reading is sent to the output buffer. For frequency
measurements, the meter uses only one "range" for all inputs between 3Hz and
300kHz. With no input signal applied, frequency measurements return "0".