Reading Memory (Store & Recall) - Picotest M3500A User Manual

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

button then use
and
buttons to find SYSTEM menu. Press ENTER
to choose it, and then again use
and
buttons to find BEEP menu.
Press ENTER to choose it, and then find ON or OFF by using
and
buttons. Press ENTER on your selection. The locations of the buttons are
shown with red rectangle frames in Figure 4-27.
Procedures: MENU → → → → SYSTEM → → → → BEEP→ → → → {ON|OFF}
Figure 4-27
Remote Interface Operation
The following commands show how to use the remote interface operation
to disable or enable the beeper:
SYSTem:BEEPer
SYSTem:BEEPer:STATe {OFF|ON}
4.4.3 Reading Memory (Store & Recall)
M3500A has a memory capacity of 2000 readings. The readings are
stored in first-in-first-out order and the memory type is volatile, which
means the stored readings will be cleared when the multimeter is
power-off. The reading memory feature can be used for all measurement
functions, math operations, trigger operations and reading hold.
How to use the reading memory
You can store the readings and access to the stored readings through
either the front panel operation or the remote interface operation.
Note: Each datum stored from M3500A to remote interface will be in
97 97 97 97