The Math Operation Command - Picotest M3500A User Manual

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

ROUTe:SCAN:COUNT <value>
Set the number of times of scanning.
ROUTe:SCAN:STATe?
Query a channel numeric which is being scanned.
ROUTe:SCAN:SCAN
Run SCAN mode
ROUTe:SCAN:STEP
Run STEP mode
C.6 The Math Operation Command
CALCulate:FUNCtion {PERCent|AVERage|NULL|LIMit|MXB|DB|DBM}
Select the math function. Only one function can be enabled at a time. The
default function is percent.
CALCulate:FUNCtion?
Query the present math function. Returns PERC, AVER, NULL, LIM, MXB, DB or
DBM.
CALCulate:STATe {OFF|ON}
Disable or enable the selected math function.
CALCulate:STATe?
Query the state of the math function. Returns "0"(OFF) or "1"(ON).
CALCulate:PERCent:TARGet {<value>|MINimum|MAXimum}
Set the target value for percent math function. The multimeter clears the value
when Min/Max is turned on, when the power has been off or a remote interface
reset.
CALCulate:PERCent:TARGet? [MINimum|MAXimum]
Query the target value for percent math function.
156
156
156
156