The System-Related Commands - Picotest M3500A User Manual

6.5 digit
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

TRIGger:DELay?
Query the trigger delay time.
TRIGger:DELay:AUTO {OFF|ON}
Disable or enable a automatic trigger delay. The delay is determined by function,
range, integration time, and ac filter setting. Specifying a delay time
automatically turns off the automatic trigger delay.
TRIGger:DELay:AUTO?
Query the automatic trigger delay mode. Returns "0" (OFF) or "1" (ON).
SAMPle:COUNt {<value>|MINimum|MAXimum}
Set the number of readings (samples) the multimeter takes per trigger. Select a
number from 1 to 50,000 readings per trigger.
SAMPle:COUNt ? [MINimum|MAXimum]
Query the sample count.
TRIGger:COUNt {<value>|MINimum|MAXimum|INFinite}
Specify the number of triggers the multimeter will accept before returning to the
"idle" state. Select from 1 to 50,000 triggers. The INFinite parameter instructs
the multimeter to take triggers continuously. Trigger count is ignored while in
local operation.
TRIGger:COUNt? [MINimum|MAXimum|INFinite]
Query the trigger count. Returns "9.90000000E+37" when you select the
infinite trigger count.
C.8 The System-Related Commands
FETCh?
Transfer readings stored in memory by the INITiate command to output buffer
where you are able to read them into your bus controller.
READ?
Change the state of the triggering system from the "idle" state to
"wait-for-trigger" state. The meter will start to make measurements when a
160
160
160
160