Status Reporting Commands - Picotest M3510A User Manual

6 1/2
Table of Contents

Advertisement

9.
Ohms Overload
10.
Not Used
11.
Limit Failed at LO
12.
Limit Failed at HI
13.
Not Used
14.
Not Used
15.
Not Used
The following conditions will clear the questionable data event
register.
1.
Users execute a *CLS command.
2.
Users query the event register by using
STATus:QUEStionable:EVENt?.
The following conditions will clear the questionable data enable
register.
1.
Users turn on the power without using *PSC command.
2.
Users execute the STATus:PRESet command.
3.
Users execute the STATus:QUEStionable:ENABle 0 command.
C.10 Status Reporting Commands
SYSTem:ERRor?
Query the multimeter's error queue. Up to 20 errors can be stored in the
queue. Errors are retrieved in first-in-first out (FIFO) order. Each error
string may contain up to 80 characters.
STATus:QUEStionable:ENABle <enable value>
Enable bits in the Questionable Data enable register. The selected bits are then
reported to the Status Byte.
STATus:QUEStionable:ENABle?
Query the Questionable Data enable register. The multimeter returns a
binary-weighted decimal representing the bits set in the enable register.
STATus:QUEStionable:EVENt?
Query the Questionable Data event register. The multimeter returns a
512
Range overload on 2-/4-wired ohm function.
1024
Set to 0
Reading is less than lower limit under limit test.
2048
4096
Reading is excess upper limit under limit test.
8192
Set to 0
Set to 0
16384
32768
Set to 0
137
137
137
137