Ph4 R Magazine Door Sensor - Kodak Ektascan 1120 Service Manual

Hide thumbs Also See for Ektascan 1120:
Table of Contents

Advertisement

SERVICE MANUAL

PH4 R Magazine Door Sensor

Note
This procedure cannot be done on machines after S/N 500249. DETECTORS were changed from
PHOTODETECTORS to MICROSWITCHES to comply with European standards.
To Check:
[1] Install the PORTABLE COMPUTER. See "Using the Portable Computer" in the Diagnostic Section.
[2] Remove the RIGHT IMAGE UNIT COVER. If necessary, see the procedure on Page 2–3.
MAIN MENU
1 Monitor Mode
4 System Params
2 System Logs
5 Debug Utility
3 Boards Diag
6 Exit
Enter Menu Item:
MONITOR MODE MENU
1 Motors
6 Seq Board Diag
2 Solenoids
7 Enter Commands
3 Relay
8 Set Beam-Splitter
4 Miscellaneous
9 Auto Sensor Test
5 Sensors
10 Semi-Auto Sensor Test
Enter Menu Item:
[6] On the PORTABLE COMPUTER, check that the R MAGAZINE DOOR SENSOR PH4 reads 0 when the
DOOR is closed, and 1 when the DOOR is open.
To Adjust:
[7] Loosen the 2 SCREWS.
[8] Move the SENSOR BRACKET until the SENSOR reads 0 when the DOOR is closed and 1 when the DOOR
is open.
[9] Tighten the 2 SCREWS.
SCREW (2)
5–90
R MAGAZINE
SENSOR
DOOR SENSOR
BRACKET
[3] Use the PORTABLE COMPUTER to display the
CES Main Menu.
[4] Enter [1] for "Monitor Mode".
[5] Enter [5] for "Sensors".
H129_1908BCC
H129_1908BC
10FEB97 – SM3226-1