Tone Scaling - Kodak Ektascan 1120 Service Manual

Hide thumbs Also See for Ektascan 1120:
Table of Contents

Advertisement

Tone Scaling

Important
• Before doing a Tone Scale adjustment, check that the Imaging Device Monitor has been correctly adjusted. This
Monitor adjustment should not be changed once the Tone Scaling has been completed.
• Tone Scale adjustments should only be done by trained personnel.
Tone Scaling is a technique of making printed images match the images on the input monitor. To adjust the tone
scale, use the following procedure:
[1] Check that the Density is set for the desired maximum density using the Setup Procedure beginning on
Page 4–22.
[2] Calibrate the PRINTER using the Calibration Procedure on Page 4–4.
[3] Use the Keypad [Program] key to adjust the Curve Shape. If necessary, see the User Manual, 5B9622.
[4] Use the Keypad [Program] key to adjust the Contrast. If necessary, see the User Manual, 5B9622.
SM3226-1 – 10FEB97
Operation Adjustments
4–13