error=0
Enter Command: PMS
Pulse number? (-30000 to 30000)[0]: -64
Speed? (10 to 1500 pps) [10]: 300
Enter Command: PMS
How many pulses? (-30000 - 30000):
How many pulses? (-30000 - 30000): -10
Speed? (10 - 1500 pps):
Speed? (10 - 1500 pps): 10
error=0
Enter command:
5–96
[25] Enter [P] [M] [S].
[26] Enter [-] [6] [4] for the pulse number.
[27] Enter [3] [0] [0] for the speed.
[28] Loosen the 2 SET SCREWS on the SUPPLY
ROLLER POSITION SENSOR FLAG.
Note
The LED indicator on the underside of the sensor is on
when the sensor is not blocked, and off when the sensor
is blocked. The flag should be set at the point just
before the LED indicator goes off. The LED indicator
should be on.
[29] Rotate the FLAG counterclockwise so that the
flag edge is located where it just blocks the sensor
beam.
[30] Check the LED indicator on the underside of the
sensor and rotate the flag clockwise to the position
just after the LED goes on.
[31] Carefully tighten the 2 SET SCREWS.
[32] Enter [P] [M] [S] to start the SUPPLY ROLLER
MOTOR.
[33] Enter [-] [1] [0]for the number of pulses.
[34] Enter [1] [0] for the speed.
[35] Check that the LED indicator on the underside of
the SUPPLY ROLLER POSITION SENSOR is
off. If the LED indicator is off, go to Step 8. If
the LED indicator is on, go to Step 21.