Measuring And Entering Density Data; Using A Stand-Alone Densitometer - Kodak Ektascan 1120 Service Manual

Hide thumbs Also See for Ektascan 1120:
Table of Contents

Advertisement

SERVICE MANUAL

Measuring and Entering Density Data

Using a Stand-alone DENSITOMETER

Old-Type Calibration Image
D-MAX
New-Type Calibration Image
D-MAX
CALIBRATION IMAGE
H129_0107AC
4–6
Use the following method to measure and enter the
density data.
[1] Prepare a stand-alone DENSITOMETER for use.
[2] Measure the density at the center of each gray step
on the Calibration Image.
(a) Check that the notch on the edge of the film
(b) Read the 9 gray steps on the calibration
D-MAX
If you have an "old-type" calibration image, read either
side strip of 9 gray steps, but read all the steps from the
H129_0101ACC
same side.
H129_0101AA
[3] Record the data.
[4] Press [SELECT] on the CONTROL PANEL to
display the screen for the first gray step value.
[5] To change the density value for a step,
(a) press the [
(b) press the [
The minimum density value that can be entered is 0.00,
and the maximum density is 3.99.
H129_0107ACA
[6] Press [NEXT] to enter the density value and
display the next density step.
[7] Do Steps 5 and 6 for gray steps 2 through 9,
sequentially.
is at the upper right corner for 35 x 43 cm, 35
x 35 cm, and 8 x 10 in. For 11 x 14 in., the
notch should be at the lower right corner.
image.
Note
ENTER STEP 1: x.xx
] key on the CONTROL
PANEL to decrease the density value.
] key on the CONTROL
PANEL to increase the density value, or
Note
10FEB97 – SM3226-1