Phf Film Suction Sensor - Kodak Ektascan 1120 Service Manual

Hide thumbs Also See for Ektascan 1120:
Table of Contents

Advertisement

SERVICE MANUAL

PHF Film Suction Sensor

Note
Use the PORTABLE COMPUTER to check if the sensor is malfunctioning before doing this procedure. If necessary,
see the Diagnostics Manual DG3226.
MAIN MENU
1 Monitor Mode
4 System Params
2 System Logs
5 Debug Utility
3 Boards Diag
6 Exit
Enter Menu Item:
MONITOR MODE MENU
1 Motors
6 Seq Board Diag
2 Solenoids
7 Enter Commands
3 Relay
8 Set Beam-Splitter
4 Miscellaneous
9 Auto Sensor Test
5 Sensors
10 Semi-Auto Sensor Test
Enter Menu Item:
Input 'HLP' to know commands.
Enter command:
Enter Command: VPO
error=0
Enter Command: VPF
error=0
5–74
To Check:
[1] Install the PORTABLE COMPUTER. See
"Using the Portable Computer" in the Diagnostic
Section.
[2] Use the PORTABLE COMPUTER to display the
CES Main Menu.
[3] Enter [1] for "Monitor Mode".
[4] Enter [7] for "Enter Commands".
Note
• Commands must be entered using capital letters.
• See the Diagnostics section to clear any error
messages displayed on the PORTABLE
COMPUTER during this procedure.
[5] Remove the RIGHT IMAGE UNIT COVER. If
necessary, see the procedure on Page 2–3.
[6] Measure the distance between the FLAG and the
edge of the SENSOR. The distance should be 1
mm (.04 in.).
[7] Enter [V] [P] [O] to start the
ELECTROMAGNETIC PUMP.
[8] Install a PLASTIC RULER or small piece of film
under the SUCKER PADS.
[9] Measure the distance between the FLAG and the
edge of the SENSOR. The distance should be 2
mm (.08 in.).
[10] Enter [V] [P] [F] to turn off the
ELECTROMAGNETIC PUMP.
[11] If necessary, do the adjustment procedure.
10FEB97 – SM3226-1