Adjustment of the Right and Left Film Guides

If there is damage to the TABS that are located on the bottom of the MAGAZINES, adjust the RIGHT and LEFT
MAGAZINE GUIDES ont he SUPPLY or RECEIVE DOOR. Use the following specifications:
Dimension
Note
It may be easier to make a gauge from a used film that is cut to the above dimensions.
MAGAZINE
OPENING
4–24
Receive Door
A
56.8 to 57.2 mm
(2.24 to 2.25 in.)
B
367.0 to 368.0 mm
(14.45 to 14.49 in.)
LEFT
MAGAZINE
GUIDE
(16.42 to 16.46 in.)
B
SCREW (3)
Supply Door
57.8 to 58.2 mm
(2.28 to 2.29 in.)
417.0 to 418.0 mm
A
RIGHT
MAGAZINE
GUIDE
MAGAZINE
TAB (4)
10FEB97 – SM3226-1