Kodak Ektascan 1120 Service Manual page 71

Hide thumbs Also See for Ektascan 1120:
Table of Contents

Advertisement

Old pulse number: XXX
Pulse number? (0 - 200): (table number)
Pulse number? (0 - 200):
Density 0=XXX pulses
error=0
Enter Command:
SM3226-1 – 10FEB97
ALIGNMENT
VALUE LABEL
H129_0708ACA
H129_0708AC
[15] Compare the pulse count values displayed on the
PORTABLE COMPUTER with the values on the
ALIGNMENT VALUE LABEL.
(a) Check the old pulse number for density
setting 0 displayed on the PORTABLE
COMPUTER.
(b) Check the pulse number for the density
setting 0 value on the ALIGNMENT
VALUE LABEL located on the front of the
OPTICAL UNIT.
[16] If the pulse number for density setting 0 value
displayed on the PORTABLE COMPUTER is
different than the setting 0 value on the
ALIGNMENT VALUE LABEL:
(a) record the value displayed on the
PORTABLE COMPUTER to be re-entered
later
(b) enter the label value at the "How many
pulses?" prompt and press the [ENTER]key
again.
Diagnostic Check Procedures
3–7