Kodak Ektascan 1120 Service Manual page 191

Hide thumbs Also See for Ektascan 1120:
Table of Contents

Advertisement

Enter Command: RMC
Check the Sucker's positions:
Are you ready?,Y/N (0/1) [0]: 0
error=0
Enter Command:
[10] Remove the RECEIVE MAGAZINE OPEN/CLOSE ASSEMBLY. If necessary, see the procedure on
Page 2-50.
[11] Check that the distance between the SPRING and the SLIDE MECHANISM is 1mm (0.04 in.) to 2mm (0.08
in.) deflection with the SLIDE MECHANISM fully extended against the SPRING.
1 mm to 2 mm
SPRING
To Adjust:
[12] Loosen the SCREW.
[13] Move the SENSOR BRACKET until the distance is correct.
[14] Tighten the SCREW.
[15] Check that the adjustment is correct.
SM3226-1 – 10FEB97
SLIDE MECHANISM
[9] Enter [R] [M] [C] to close the RECEIVE
MAGAZINE WINDOW.
SENSOR
BRACKET
SCREW
Electrical Adjustments
R MAGAZINE
CLOSE SENSOR
H129_0020BCA
H129_0020BC
5–89