Display
Error code
Other error code

19-3. Other Display

Advise

19-4. Asking For Repair

If the balance needs service or repair, contact your local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and
observe the following when transporting the balance.
Use the original packing material for transportation.
Remove the draft gate, weighing pan and pan support from the main unit before placing the
balance in the shipping container.
Description
Timeout error
If the timeout parameter is set to "
EC,E03
receive the next character of a command within the time limit
of one second. Confirm the communication.
Excess characters error
EC,E04
The balance received excessive characters in a command.
Confirm the command.
Format error
A command includes incorrect data.
EC,E06
Example:
The data is numerically incorrect.
Confirm the command.
Parameter setting error
EC,E07
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.
If an error described above can not be released or other
errors are displayed, contact the local A&D dealer.
When this indicator blinks, automatic self calibration is required. The
indicator blinks when the balance detects changes in ambient
temperature. If the balance is not used for several minutes with this
indicator blinking, the balance performs automatic self calibration. The
blinking period depends on the operating environment.
The balance can be used while this indicator is blinking. We
recommend that you perform automatic self calibration for precision
weighing.
90
", the balance did not