Calibration report using an external weight
Setting of "info 1"
AD-8121 printer format
-----------A-&-D
MODEL-----GX-30K
S/N-----01234567
ID-------ABCDEFG
DATE--2004/12/31
TIME----12:34:56
CALIBRATED(EXT.)
CAL.WEIGHT
-----+20000.0--g
SIGNATURE
---------------
Space, ASCII 20h
Terminator, CR , LF or CR
<TERM>
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Calibration test report using an external weight
Note Calibration test does not perform calibration.
Setting of "info 1"
AD-8121 printer format
-----------A-&-D
MODEL-----GX-30K
S/N-----01234567
ID-------ABCDEFG
DATE--2004/12/31
TIME----12:34:56
CAL.TEST(EXT.)
ACTUAL
----------0.0--g
-----+19999.9--g
TARGET
-----+20000.0--g
SIGNATURE
---------------
Space, ASCII 20h
Terminator, CR , LF or CR
<TERM>
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration weight
Signature
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
54
Setting of "info 2"
General format
~~~~~~~~~~~A~&~D<TERM>
MODEL~~~~~GX-30K<TERM>
S/N~~~~~01234567<TERM>
ID~~~~~~~ABCDEFG<TERM>
DATE<TERM>
~~~~~~2004/12/31<TERM>
TIME<TERM>
~~~~~~~~12:34:56<TERM>
CALIBRATED(EXT.)<TERM>
CAL.WEIGHT<TERM>
~~~~~+20000.0~~g<TERM>
SIGNATURE<TERM>
<TERM>
<TERM>
----------------<TERM>
<TERM>
<TERM>
Setting of "info 2"
General format
~~~~~~~~~~~A~&~D<TERM>
MODEL~~~~~GX-30K<TERM>
S/N~~~~~01234567<TERM>
ID~~~~~~~ABCDEFG<TERM>
DATE<TERM>
~~~~~~2004/12/31<TERM>
TIME<TERM>
~~~~~~~~12:34:56<TERM>
CAL.TEST(EXT.) <TERM>
ACTUAL<TERM>
~~~~~~~~~~0.0~~g<TERM>
~~~~~+19999.9~~g<TERM>
TARGET<TERM>
~~~~~+20000.0~~g<TERM>
SIGNATURE<TERM>
<TERM>
<TERM>
----------------<TERM>
<TERM>
<TERM>