A&D GF-30K Instruction Manual page 67

Gx-k / gf-k series high resolution industrial balance
Table of Contents

Advertisement

Notes
Using the "CN:mm" command, the comparator limit values can be recalled.
The upper limit value recalled can be output using the "?HI" command.
The lower limit value recalled can be output using the "?LO" command.
The upper limit value can be changed using the "HI:" command.
The lower limit value can be changed using the "LO:" command.
"mm" indicates a two-digit numerical value 01 to 20, which corresponds to C01 - C20.
Digital input mode
In the digital input mode, the upper and lower limit values are
entered digitally using the keys. Use the following keys to store
upper and lower limit values in digital input mode.
key ... To select the digit to be changed.
SAMPLE
key .. To change the value of the selected digit.
RE-ZERO
key ...... To switch the polarity.
MODE
key ....... To store the upper and lower limit values.
PRINT
key .......... To return to the upper and lower limit
CAL
key
MODE
Weighing input mode
In the weighing input mode, a sample is placed on the pan to store the upper and lower limit
values.
Notes
Pressing the
balance will return to the upper and lower limit values
confirmation mode (step 3 in "Storing the upper and lower
limit values").
To go to the digital input mode, press and hold the
Step 1 The first display in the weighing input mode depends on the
comparator number selected in step 4 in "Storing the upper and
lower limit values".
For example, when "
the current weight value and the comparator number with
blinking.
Step 2 Place a container on the weighing pan, if necessary.
Press the
Step 3 Place a sample corresponding to the upper limit value,
on the pan or in the container.
Step 4 Press the
Go to step 3 of the previous page.
values confirmation mode
Go to step 3 of the previous page.
Press and hold to go to the weighing input mode.
key will interrupt the operation and the
CAL
" is selected in step 4, the display is
C03 H1
key to set the display to zero.
RE-ZERO
key to store the upper limit value.
PRINT
MODE
65
Value
Digit
Store
Polarity
C03
Upper limit value
HI
Comparator
number
key.
Place a sample
corresponding to the
upper limit value
HI
Upper
limit
Store