Data Memory; Notes On Using Data Memory - A&D GF-30K Instruction Manual

Gx-k / gf-k series high resolution industrial balance
Table of Contents

Advertisement

12. Data Memory

Data memory is a function to store weighing data, calibration data and unit mass in memory.
Of the data in memory, the balance can only display the weighing data. The weighing data
and calibration data in memory are available for outputting at one time to a printer or
personal computer. One of the following data sets can be stored:
Weighing data
Calibration result of Internal and external calibration
and calibration test
Unit mass in the counting mode
Upper and lower limit values of comparator function
Tare value
Note GF-K series does not output time and date.

12-1. Notes on Using Data Memory

To use the memory function, set the "Data memory (
In addition, for weighing data, set the "Time/Date output (
setting the data memory, refer to "10-2. The Function Table".
For weighing data of the GX-K series, the data contents to be stored and the storage
capacity depend on the "Time/Date output (
Releasing "Clr"
If a different type of data exists in memory when the data is
stored, "
" blinks the upper left of the display. For example,
Clr
you want to store weighing data but calibration data or unit mass
data remains in memory.
Under such a condition, before storing data, delete the data in memory as follows:
Step 1 Press and hold the
then release the key.
Step 2 Press the
The type of data stored in memory appears in the upper left of the display as
shown below:
Unit mass in the counting mode
Weighing data without time and date
Weighing data with time and date
Calibration result
parameters of comparator
Tare value
Step 3 Press the
Step 4 The balance displays
Excluding date and time
Including date and time
key until
PRINT
key to display
RE-ZERO
key to delete all the data in memory.
PRINT
and returns to the weighing mode.
end
GX-K series
Up to 200 sets
Up to 100 sets
Last 50 sets
Up to 50 sets
)" parameter of the function table.
data
5-td
)" parameter setting.
5-td
with "no" blinking is displayed,
SClr " n " o
with "go" blinking.
SClr " g " o
PpC
-d-
d-t
Hi5
Cp
tr
56
GF-K series
Up to 40 sets
-
-
Up to 20 sets
Up to 20 sets
Up to 20 sets
)" parameter. For details on
Upper left of
the display

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents