Percent Mode (%); Animal Weighing Mode (Hold Function) - A&D GF-30K Instruction Manual

Gx-k / gf-k series high resolution industrial balance
Table of Contents

Advertisement

6-3. Percent Mode (%)

The percent mode displays the weighting value in percentage compared to a 100%
reference mass and is used for target weighing or checking the sample variance.
Selecting The Percent Mode
Step 1 Press the
(Percent mode). If the percent mode can not be
selected, refer to "5. Weighing Units".
Storing The 100% Reference Mass
Step 2 Press the
reference mass storing mode.
Even in the storing mode, pressing the
key will switch to the next mode.
Step 3 Place a container on the weighing pan, if
necessary. Press the
weight (tare). The balance displays
Step 4 Place the sample to be set as the 100%
reference mass on the pan or in the container.
Step 5 Press the PRINT key to store the reference
mass. The balance displays
decimal point position depends on the reference
value. The reference mass stored, even if the
AC adapter is removed, is maintained in
non-volatile memory.)
Note
If the balance judges that the mass of the
sample is too light to be used as a reference, it
displays
Step 6 Remove the sample.
Reading The Percentage
Step 7 Place a sample to be compared to the reference
mass on the pan. The displayed percentage is
based on the 100% reference mass.

6-4. Animal Weighing Mode (Hold Function)

This is the mode to weigh a moving object such as an animal, even when the display of the
weighing data fluctuates. The hold function allows the average weight of the animal to be
displayed. To use the hold function, set the function in the function table. Refer to "10.
Function Table" and "10-3. Description Of The Class "Environment, Display" " for details.
key to select the unit
MODE
key to enter the 100%
SAMPLE
RE-ZERO
.
lo
%
MODE
key to cancel the
.
100 0 %
. (The
100. 0 0 %
19
Pan
Container
100%
mass
Place sample
Percentage