Other Display

19.3.
When this indicator blinks, automatic self calibration is required. The indicator
blinks when the balance detects changes in ambient temperature. If the balance
is not used for several minutes with this indicator blinking, the balance performs
automatic self calibration. The blinking duration depends on the operating
environment.
Note The balance can be used while this indicator is blinking. We recommend that you
perform automatic self calibration for precision weighing.
This is the display to prompt you to enter a password. When the password function is
used, the correct password must be entered for the balance to enter the weighing
mode.
For details, refer to "15. Password Function".
BM series
Not ready
A received command cannot be processed.
Examples:
The balance received a "Q" command, but not in the weighing mode.
The balance received a "Q" command while processing a RE-ZERO
command.
Adjust the delay time to transmit a command.
Timeout error
When the timeout parameter is set to "
receive the next character of a command within the time limit of one
second. Confirm the communication.
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
Format error
A command includes incorrect data.
Example:
The data is numerically incorrect.
Confirm the command.
Parameter setting error
The received data exceeds the range that the balance can accept.
Confirm the parameter range of the command.
If an error described above cannot be resolved or other errors are
displayed, contact the local A&D dealer.