Control Using Cts And Rts; Settings Related To Rs-232C - A&D GF-30K Instruction Manual

Gx-k / gf-k series high resolution industrial balance
Table of Contents

Advertisement

17-3. Control Using CTS And RTS

Depending on the "Ct5" parameter of "Serial interface (5if)", the balance performs as follows:
Ct5 0
Regardless of whether the balance can receive a command or not, the balance keeps the
CTS line HI. The balance outputs data regardless of the condition of the RTS line.
Ct5 1
The CTS line is kept Hi normally. When the balance can not receive the next command
(Example: while the balance is processing the last command), the balance sets the CTS line
to Lo. The balance confirms the level of the RTS line before outputting a set of data. If the
RTS level is Hi, the balance outputs data. If the RTS level is Lo, data is not output (The data
is canceled).

17-4. Settings Related To RS-232C

Concerning the RS-232C, the balance has two functions: "Data output (
interface (
)". Set each function as necessary.
5if
85
)" and "Serial
dout