Connection To Peripheral Equipment - A&D GF-30K Instruction Manual

Gx-k / gf-k series high resolution industrial balance
Table of Contents

Advertisement

Connection to peripheral equipment

16-2.
16-2-1.
Connection To The AD-8121B Printer
Preset the following parameters to use the AD-8121B printer.
Item and Parameter Factory
Class
prt
Data output mode
ap-p
Auto print polarity
ap-b
Auto print difference
d-no
dout
Data number output
Data output
5-td
Time/Date output
5-id
ID number output
pU5e
Data output pause
at-f
Auto feed
bp5
Baud rate
btpr
Data bit, parity bit
5if
Crlf
Serial
Terminator
interface
type
Data format
Ct5
CTS, RTS control
#1 Set parameters when auto print mode A or B (prt 1 or 2) is selected.
#2 Set 1 when multiple lines are printed. Example: When appending ID number, set 1.
Settings of AD-8121B
AD-8121B
MODE
DIP switch
MODE 1
MODE 2
MODE 3
DIP switch No.3 : Handling unstable data
ON
Print
OFF
Not printed
DIP switch No.4 : Data input specifications (Interface selection)
ON
Current loop
OFF
RS-232C
The printer performs as follows, depending on the data memory setting.
Setting
Output data
The weighing data
data 0
The weighing data stored in memory
data 2
The calibration report stored in memory
data 3
Refer to "11-2. GLP Report" for print samples.
GF-K series does not output the time and date. Use the calendar function of the AD-8121B.
GF-K series does not store the calibration report in memory.
MODE
Print at receiving data.
4
3
2
1
Standard mode, statistic mode
MODE
Print by DATA key operation or built-in timer.
4
3
2
1
Standard mode, interval mode, chart mode
MODE
Print at receiving data.
4
3
2
1
Dump print mode
AD-8121B
settings
MODE 1
0
0,1,2
0
#1
1
0
0
0
0
0
0
0
0
0
0
2
2
0
0
0
0
0
0
0
0
Description
MODE
4
3
2
1
MODE
4
3
2
1
81
AD-8121B
AD-8121B
MODE 2
MODE 3
3
0,1,2
Not
#1
necessary
0
0,1
0
0,1,2,3
0
0,1
0
0,1 #2
0
0,1
2
2
0
0
0
0
0
1
0
0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents