Ac:point; Data:ac; Data[:Dc] - HP E1340A User Manual

Arbitrary function generator
Hide thumbs Also See for E1340A:
Table of Contents

Advertisement

:AC:POINt

Comments

:DATA:AC

Comments

:DATA[:DC]

Comments
Chapter 7
CALibration:AC:POINt <
AC calibration point, computes needed calibration constants, and sets up the
HP E1340A for the next measurement. Each calibration constant is checked
for validity when it is computed. When all measurements and computations
are completed, the constants are stored in the HP E1340A's non-volatile
calibration memory.
The
command should be sent after completing the calibration
*RST
procedure to restore normal operation.
See the " Calibration" section of the HP E1340A Service Manual for
detailed information on the use of this command.
Most of the HP E1340A's commands cannot be executed while
calibration is in progress. The
prematurely terminate the calibration procedure without affecting the
stored calibration constants.
Executable when initiated: No
Coupled command: No
Related commands:
*RST Condition: none
CALibration:DATA:AC <
E1340A's calibration constants in IEEE-488.2 arbitrary block program data
format. The query form returns this portion of the calibration constants in
IEEE-488.2 definite block data format. See the " Calibration" section of the
HP E1340A Service Manual for detailed information on the use of this
command.
Executable when initiated: No
Coupled command: No
Related commands: none
*RST Condition: unaffected
CALibration:DATA[:DC] <
E1340A's calibration constants in IEEE-488.2 arbitrary block program data
format. The query form returns the current DC portion of the calibration
constants in IEEE-488.2 definite block data format. See the " Calibration"
section of the HP E1340A Service Manual for detailed information on the
use of this command.
Executable when initiated: No
Coupled command: No
Related commands: none
*RST Condition: unaffected
value
>
takes the measured value for the current
command may be used to
*RST
CALibration:AC:BEGin
block
transfers the AC portion of the HP
>
block
>
transfers the DC portion of the HP
HP E1340A Command Reference
CALibration:AC:POINt
143