Calibrating the Network Analyzer
Performing the Calibration
Performing the Calibration
Step 1. Make the connections shown in
The figure shows two RF cables connecting the two ports of the network
analyzer to the two ports of the ECal Module. The 24" RF cables
(Agilent Part No. 8120-4781 or equivalent) listed in
page 2-3
but not overtorqued. A torque wrench set to 12 inch-pounds (Agilent
Part No. 8710-1766 or equivalent) is recommended.
Figure 4-1
Calibrating the Network Analyzer
To USB Port
on Rear Panel
4-2
should be used. Make sure the connectors are snugly tightened
USB Quad Hub
Port 1 Port 2 Port 3 Port 4
Up Link
USB Cable
Keyboard with USB Cable
Port 1
Figure
4-1.
Mouse with
USB Cable
Network Analyzer
USB Port
Port 2
RF Cables
Electronic
Calibration
Module
Agilent Technologies Z5623A Option H48
Table 2-2
on
USB Cable