Performance Verification
Setting Up Limit Testing
Step 5. The Scale window shown in
the following values:
• Scale Per Division = 5 dB
• Reference Level = 0 dB
• Reference Position = 9 Divisions
When these values have been entered, the Scale window will look like
Figure
Figure 5-20
The Scale Window With Desired Values For Return Loss

Setting Up the Limit Table

Step 1. Click Trace on the menu bar. When the Trace menu appears, scroll
down to the Limit Test... button and click it. See
Step 2. When the Limit Test window pops up as shown in
Show Table button, then click the OK button.
Step 3. When the Limit Table pops up as shown in
from
shortcuts listed on
Table 5-5
Limit Table Entries for Return Loss (Port Active)
Segment
Number
5 -20
5-20. Click the OK button.
Table 5-5
to the Limit Table on the network analyzer. (Consult the
page
5-13.) The result is shown in
Segment
Beginning
Type
Stimulus
1
Max
300 kHz
2
Max
1.3 GHz
3
Max
4
Max
Figure 5-6
will pop up. When it does, set
Figure
Ending
Stimulus
1.3 GHz
3 GHz
3 GHz
6 GHz
6 GHz
9 GHz
Agilent Technologies Z5623A Option H48
Figure
5-8.
Figure
5-9, click the
5-10, transfer entries
Figure
5-21.
Beginning
Ending
Response
Response
−25.5 dB
−25.5 dB
−19 dB
−19 dB
−14.5 dB
−14.5 dB
−9.5 dB
−9.5 dB