Specifying If Bandwidth And Averaging - Agilent Technologies Option H48 Multiport Test Set Z5623A User's And Service Manual

Multiport test set
Table of Contents

Advertisement

Performance Verification
Setting Up Limit Testing

Specifying IF Bandwidth and Averaging

Step 1. Click Sweep on the menu bar. When the Sweep menu appears, scroll
down to the IF Bandwidth... button and click it. See
Figure 5-28
The "IF Bandwidth..." Button
Step 2. When the IF Bandwidth window pops up as shown in
change the IF Bandwidth to 100Hz. (Type, "100<space>", without the
quotation marks.) The result is shown in
button.
Figure 5-29
IF Bandwidth Window with Default Value
5 -28
Figure
5-30. Click the OK
Agilent Technologies Z5623A Option H48
Figure
5-28.
Figure
5-29,

Advertisement

Table of Contents
loading

This manual is also suitable for:

Z5623a option h48

Table of Contents