Step 2. Press the Preset button on the Network Analyzer.
Step 3. Verify that the Ready light on the ECal Module is green.
Step 4. On the network analyzer, click Calibration on the menu bar, then click
the Calibration Wizard button. See
Figure 4-2
Calibration Menu on Network Analyzer
Step 5. When the window shown in
button. The Cal Type window shown in
Figure 4-3
Cal Wizard Window
The currently selected calibration type is "None".
Agilent Technologies Z5623A Option H48
Calibrating the Network Analyzer
Figure
4-2.
Figure 4-3
comes up, click the Cal Type
Figure 4-4
Performing the Calibration
will then pop up.