Agilent Technologies Option H48 Multiport Test Set Z5623A User's And Service Manual page 55

Multiport test set
Table of Contents

Advertisement

Figure 4-7
Electronic Calibration Window
Step 10. Click the Module button. The window shown in
pop up.
Figure 4-8
Module Connection Message
Step 11. Click the OK button. (The connections requested in
made in
window will close, the window shown in
and the calibration will begin.
The network analyzer will complete the calibration with no further
operator actions required. The calibration will take about 5 seconds.
Step 12. Verify that the calibration has been completed. This is shown by a check
mark over the Module button as shown in
Figure 4-9
Electronic Calibration Window After Cal Is Complete
Step 13. Click the OK button shown in
Cal Wizard window will pop back up with the following message as
shown in
Agilent Technologies Z5623A Option H48
Step 1
of
"Performing the
Figure
4-10: "Done. Save your settings now by pressing OK".
Calibrating the Network Analyzer
Figure 4-8
Calibration" on
page
Figure 4-7
Figure
4-9.
Figure
4-9. The window will close; the
Performing the Calibration
will then
Figure 4-8
were
4-2.) This
will pop back up,
4-5