Step 12. Give the file a new name that includes the following:
• The date the cal was performed
• The suffix ".cst"
For example, if the calibration was performed on 6/17/01, a filename of
"cal_06_17_01.cst" would be appropriate. See
Figure 4-20
The Cal File After It Has Been Renamed
Step 13. Record the new name of the calibration file in
have a suffix of ".cst".
Table 4-1
Name of Calibration File
Agilent Technologies Z5623A Option H48
Calibrating the Network Analyzer
Renaming the Calibration File
Figure
4-20.
Table
4-1. The file must
.cst
4-13