FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
Page 3
Safety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document. Warning denotes a hazard. It calls attention to a procedure WARNING which, if not correctly performed or adhered to, could result in injury or loss of life.
This User's and Service Guide documents the use of the test set with an NOTE Agilent E8358A network analyzer. Figure 1-1 Typical Equipment Setup Port 1 Reflection Port Figure 1-1 Network Analyzer Port 2 RF Cables Z5623A Option H48 Transmission Port Test Ports Agilent Technologies Z5623A Option H48 shows a typical equipment...
Agilent provides two different types of specifications for the test set: • Guaranteed performance specs • Typical performance specs Typical performance specs have been benchmarked during product development, but are not tested by the factory and are not guaranteed. Agilent Technologies Z5623A Option H48 Instrument Description Specifications...
4) 9 dB 1) 26 dB 2) 20 dB 3) 16 dB 4) 14 dB 1) 1.5 dB 2) 2.0 dB 3) 2.5 dB 4) 3.5 dB 5 million cycles 1 Watt (RF + DC) Agilent Technologies Z5623A Option H48...
Band 4 (6.0 GHz to 9.0 GHz) f. IF Bandwidth set to 10 Hz g. Conditions: 2 Windows, 4 Channels, 8 Traces, 2-port cal, no band crossings, 201points. Agilent Technologies Z5623A Option H48 Typical Performance 1) –8.0 dBm 2) –8.0 dBm 3) –7.0 dBm...
POLLUTION DEGREE 2, per IEC 101 and 664 respectively. Non-Operating Storage Conditions Temperature: –40 °C to +70 °C Humidity: 0 to 90 percent relative at +65 °C (non-condensing) Altitude: 0 to 15,240 meters (50,000 feet) Agilent Technologies Z5623A Option H48...
Option 908, part number 5062-3974, provides rack mounts that make it quick and easy to install or remove the test set from a mainframe. For further information on these options, please contact Agilent Technologies. Refer to Chapter 10 , “Contacting Agilent.” Agilent Technologies Z5623A Option H48...
Cleaning and Shipping Instructions Cleaning and Shipping Instructions Cleaning Clean the instrument cabinet using a damp cloth only. Shipping Always transport or ship the instrument using the original packaging if possible. If not, comparable packaging must be used. Agilent Technologies Z5623A Option H48...
Recommended Additional Equipment • Familiarization with Safety Requirements • Electrical Preparations • Environmental Preparations • Test Set Familiarization • Setting the GPIB Address of the Test Set • Connecting and Turning on the Test Set Agilent Technologies Z5623A Option H48 2 -1...
Inspect the test set and all accessories for any signs of damage that may have occurred during shipment. If your test set or any accessories appear to be damaged or missing, call Agilent Technologies. Refer to Chapter 10 , “Contacting Agilent.”...
Type N 50 Ohm Female-Female Adapter a. Or equivalent b. Available from Peracom Networks, Inc., Cary, North Carolina (www.peracom.com). Agilent Technologies Z5623A Option H48 Recommended Additional Equipment Table 2-2 is not shipped with the test set, but is Model Or Part Number Agilent Model No.
3. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply. 2 -4 are met. to see the different types of power Figure 2-2, call Chapter 10 , “Contacting Agilent.” Agilent Technologies Z5623A Option H48...
Page 21
Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption of the protective conductor is prohibited. Agilent Technologies Z5623A Option H48 Installation Electrical Preparations 2 -5...
8120-4600 8120-5182 250V 8120-5181 a. E =earth ground, L = line, and N = neutral. b. Plug identifier numbers describe the plug only. The Agilent Technologies part number is for the complete cable assembly. 2 -6 Plug Length Cable Description cm (in.)
Figure 2-3 Ventilation Clearance Requirements If the total power dissipated in the cabinet is greater than 800 watts, CAUTION forced convection must be used. Agilent Technologies Z5623A Option H48 “Environmental Requirements” on page 1-6 Figure 2-3. The convection into and Installation Environmental Preparations are met.
Page 24
Environmental Preparations 3. Set up a static safe workstation such as the one shown in Electrostatic discharge (ESD) can damage or destroy electronic components. Figure 2-4 Example of an Antistatic Workstation 2 -8 Figure Agilent Technologies Z5623A Option H48 2-4.
CAUTION damage to the internal RF switches or the analyzer may occur. The GROUND Connector The GROUND Connector provides a convenient front panel ground connection for a standard banana plug. Agilent Technologies Z5623A Option H48 PORT CONNECTION TRANSMISSION Status LCD...
Page 26
250 mA on any control line (pins 1 through 5 of the CONTROL LINES Connector) when that line is commanded to the “Low” state (+0.63 Vdc). 2 -10 to“Using the Control for further details. Agilent Technologies Z5623A Option H48...
Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited. Agilent Technologies Z5623A Option H48 Test Set Familiarization GPIB GPIB Connector...
Page 28
The line fuse (F 3 A/250 V, part number 2110-0780) and a spare reside within the line module. Figure 2-7 illustrates where the fuses are and how to access them. Figure 2-7 Location of Line Fuses 2 -12 Agilent Technologies Z5623A Option H48...
(16 + 8 + 4 + 2 + 1 = 31.) Therefore, the GPIB address can have any value from 0 to 31 inclusive. Agilent Technologies Z5623A Option H48 2-11. A diagram of the switch bank is presented 2-8.
2 -14 Figure 2-9, “Detailed Equipment The figure shows two RF cables connecting the Part No. Purpose 8120-4781 Verification of test set performance 8120-4782 Measurement of multiport devices Agilent Technologies Z5623A Option H48 Table 2-3. Otherwise, use...
Page 31
To GPIB Connector on Rear Panel GPIB Cable Reflection Port To GPIB Connector on Rear Panel Agilent Technologies Z5623A Option H48 USB Quad Hub Port 1 Port 2 Port 3 Port 4 Up Link Mouse with USB Cable Keyboard with USB Cable...
Page 32
Installation Connecting and Turning on the Test Set 2 -16 Agilent Technologies Z5623A Option H48...
How to Send a Manual Command • Common Errors in Manual Commands • A List of GPIB Commands • Restoring the Analyzer to its Normal Configuration • Disaster Recovery Agilent Technologies Z5623A Option H48 Chapter 2 , “Installation,” have been 2-15.
Step 2. Make sure that both the PNA Series analyzer and the test set are turned on. Step 3. On the network analyzer, click the System menu, scroll down to Configure, and click SICL/GPIB. See Figure 3-1 System Menu Figure 2-9 on page 2-15 Figure 3-1, “System Menu.” Agilent Technologies Z5623A Option H48 have...
Page 35
When the Minimize button is visible, minimize the analyzer window by clicking on the button. Figure 3-3 Minimize Window Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set Preparing the Analyzer to Control the Test Set Figure 3-2.
Page 36
Step 7. From the Measurement & Automation window, click on the Devices and Interfaces folder to expand the directory. Step 8. Right-click GPIBO (AT-GPIB/TNT), then click Interactive Control. Figure 3-5, “Interactive Control window.” Figure 3-5 Interactive Control window Agilent Technologies Z5623A Option H48...
Page 37
Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set Preparing the Analyzer to Control the Test Set C:\Progam Files\National Instruments\ NI-488.2\...
Page 38
Step 12. Use the mouse to size and position the PNA Series Network Analyzer window and the command window ( C:\Progam Files\National windows are Instruments \NI-488.2\bin\ibic.exe) so that both visible on the network analyzer screen. See Figure 3-7, “Manual Control Window.” Figure 3-7 Manual Control Window Agilent Technologies Z5623A Option H48...
The test set then treats the string as a single command. Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set Set”) have been performed.
Three leading zeroes are missing. (This is a binary Control Line command. It needs three leading zeroes.) (Decimal Control Line commands don’t need leading zeroes.) Agilent Technologies Z5623A Option H48 Solution Send a semicolon: “;” Resend both commands with a semicolon in each: “refl_01;”...
• Reflection port of the test set. • Transmission port of the test set. • 50 Ω termination inside the test set. Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set A List of GPIB Commands...
Page 43
L = “Low”; O = “Open”; x = “Unchanged” To connect all test ports to their internal 50 Ω loads, send the following command: OUTPUT 712;"*all_term" Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set GPIB Command Decimal...
Step 2. In the analyzer window, click System on the menu bar, then scroll down to Configure, then click SICL/GPIB. Step 3. In the SICL/GPIB window, click Talker/Listener and then OK. This allows other devices to control the analyzer. 3-12 Agilent Technologies Z5623A Option H48...
X in the upper right-hand corner of the window to close the application. c. When all applications are closed, the task bar will look like the one Figure 3-9. Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set Disaster Recovery Figure 3-13...
Page 46
Using the Network Analyzer to Control the Test Set Disaster Recovery Figure 3-8 Network Analyzer Screen With Four Applications Running Figure 3-9 Network Analyzer Screen With No Applications Running 3-14 Four applications running No applications running Agilent Technologies Z5623A Option H48...
Page 47
Step 3. Click Properties in the drop-down menu. A “System Properties” window will appear as shown in Figure 3-11 “System Properties” Window Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set Disaster Recovery Figure 3-10.
Page 48
Step 6. Check to see if the “AT-GPIB/TNT” icon is visible as shown in 3-14. If it is not, click on the “National Instruments GPIB Interfaces” icon as in 3-16 Figure Figure 3-13 Figure Figure 3-13. Agilent Technologies Z5623A Option H48 3-11. The result is shown in Click this button. 3-14. Figure...
Page 49
Step 9. The following message will pop up: “Disabling this device will cause it to stop functioning. Do you really want to disable it?” Click “Yes.” See Figure 3-15. Agilent Technologies Z5623A Option H48 Using the Network Analyzer to Control the Test Set Disaster Recovery 3-17...
Next, hold the power switch in for about 20 seconds, or until the color of the switch changes from green to orange. Release the switch, then press it again and release it immediately. 3-18 “Preparing the Analyzer to 3-2. Agilent Technologies Z5623A Option H48...
Series Network Analyzer for a comprehensive tutorial on this calibration method and on many others. This chapter contains the following sections: • Performing the Calibration • Renaming the Calibration File • Recalling and Examining Calibrations Agilent Technologies Z5623A Option H48...
Port 1 Port 2 Port 3 Port 4 Up Link Mouse with USB Cable Keyboard with USB Cable Network Analyzer Port 1 RF Cables Electronic Calibration Module Agilent Technologies Z5623A Option H48 Table 2-2 USB Port Port 2 USB Cable...
Page 53
Step 5. When the window shown in button. The Cal Type window shown in Figure 4-3 Cal Wizard Window The currently selected calibration type is “None”. Agilent Technologies Z5623A Option H48 Calibrating the Network Analyzer Performing the Calibration Figure 4-2.
Page 54
Wizard window is “Full SOLT 2-Port (Omit Isolation)” as shown in Figure 4-6. Step 9. Click the Measure ECal Stds button in the Cal Wizard window. The window shown in Figure Figure 4-7 will then pop up. Agilent Technologies Z5623A Option H48 Figure 4-5. 4-6.
Page 55
Electronic Calibration Window After Cal Is Complete Step 13. Click the OK button shown in Cal Wizard window will pop back up with the following message as shown in Agilent Technologies Z5623A Option H48 “Performing the Calibration” on Figure 4-9. The window will close; the Figure 4-10: “Done.
Page 56
(The date and time can be used to locate the file in case the filename you write down is incorrect or incomplete.) Step 16 will ask you to write Step 15 above.) A typical filename Agilent Technologies Z5623A Option H48...
Step 3. The window shown in icon in that window. Figure 4-11 “My Computer” Window Agilent Technologies Z5623A Option H48 Step 5 of “Preparing the Analyzer to Control the Test Set” Figure 4-11 will pop up. Click the Hard Disk (C:)
Page 58
Figure 4-12 Root Directory of the C: Drive Step 5. The window shown in Figure 4-13 “C:\Program Files” Directory Figure 4-12 will pop up. Click the “Program Figure 4-13 will pop up. Click the “Agilent” icon. Agilent Technologies Z5623A Option H48...
Page 59
Figure 4-14 “C:\Program Files\Agilent” Directory Step 7. The window shown in icon. Figure 4-15 “C:\Program Files\Agilent\Network Analyzer” Directory Agilent Technologies Z5623A Option H48 Calibrating the Network Analyzer Renaming the Calibration File Figure 4-14 will pop up. Click the “Network Figure 4-15...
Page 60
The result is shown in Figure 4-16 Directory Containing Cal Files Figure 4-17 Directory Containing Cal Files After Clicking “Type” Button 4-10 Figure 4-16 will pop up. It contains all of the Figure Agilent Technologies Z5623A Option H48 4-17.
Page 61
(until it is highlighted as shown in Figure 4-18). Figure 4-18 Cal File Highlighted Step 10. Move the cursor off the selected filename (either to the left or to the right). The selected filename will remain highlighted. Agilent Technologies Z5623A Option H48 4-11...
Page 62
Calibrating the Network Analyzer Renaming the Calibration File Step 11. Click File on the menu bar, then scroll down and click the Rename button as shown in Figure 4-19. Figure 4-19 Preparing to Rename the Cal File 4-12 Agilent Technologies Z5623A Option H48...
Page 63
The Cal File After It Has Been Renamed Step 13. Record the new name of the calibration file in have a suffix of “.cst”. Table 4-1 Name of Calibration File Agilent Technologies Z5623A Option H48 Calibrating the Network Analyzer Renaming the Calibration File Figure 4-20.
Method 1. Calibration File Is Not On “Quick Recall” List Step 1. In the network analyzer window, click File on the menu bar, then scroll down and click the Recall button. See Figure 4-21 File Recall Button 4-14 Figure 4-21. Agilent Technologies Z5623A Option H48...
Page 65
4-23. The file will then be recalled and the window will close. Figure 4-22 File Recall Window Figure 4-23 File Recall Window After Clicking the Desired Calibration File Agilent Technologies Z5623A Option H48 Calibrating the Network Analyzer Recalling and Examining Calibrations Figure 4-22 will pop up. Move the...
To change from one state to the other, just click the button. Try it! The trace will change instantaneously. (Compare the trace in with the trace in 4-16 Figure 4-24. The file will immediately be recalled. Figure 4-26.) Agilent Technologies Z5623A Option H48 Figure Figure Figure 4-25...
Page 67
Figure 4-25 Calibration Turned On Figure 4-26 Calibration Turned Off Agilent Technologies Z5623A Option H48 Calibrating the Network Analyzer Recalling and Examining Calibrations 4-17...
Step 1. Click Calibration on the menu bar, then scroll down to the Properties button and click it. See Figure 4-27 Calibration Properties Button Step 2. The window shown in Figure 4-28 Calibration Properties Window 4-18 Figure 4-27. Figure 4-28 will pop up. Examine it carefully. Agilent Technologies Z5623A Option H48...
— Limit Testing for Return Loss (Port Active) — Limit Testing for Return Loss (Port Off) — Limit Testing for Crosstalk • Verifying Return Loss and Insertion Loss Specifications • Verifying Crosstalk Specs Agilent Technologies Z5623A Option H48 5 -1...
Page 70
Measurement Uncertainty 5-1, the plus sign is selected if the limit is a minimum Band 2 (1.3 - 3 GHz) (3 - 6 GHz) Agilent Technologies Z5623A Option H48 Table 5-1 apply to the Band 3 Band 4 (6 - 9 GHz)
Page 71
When using an external termination on the Reflection or Transmission port h.When using one of the 50 Ω internal terminations i. From any test set port to the Reflection or Transmission port Agilent Technologies Z5623A Option H48 to the uncertainties listed in Table 1-1 on page Table...
The remainder of this section describes, for each category, the S- parameter measurement that will be used to verify the specifications. The test limits of Table 5-2 will be converted to S-parameter test limits for each category. 5 -4 Agilent Technologies Z5623A Option H48...
(S because the test set provides no amplification. In units of dB, S range from zero to minus infinity.) Agilent Technologies Z5623A Option H48 Performance Verification are both in units of dB. . Limit lines Table 5-2.
−100 dB, the corresponding S below −100 dB in order for the instrument to pass. 5 -6 . By definition, limit will be set up as limit will be set up so that S Agilent Technologies Z5623A Option H48 must be...
• Limit Testing for Return Loss (Port Active) • Limit Testing for Return Loss (Port Off) • Limit Testing for Crosstalk Agilent Technologies Z5623A Option H48 page 5-4. As you work through this section, record the Table 5-3. File Name...
Step 1. Click Trace on the menu bar. When the Trace menu appears, scroll down to the Measure button and pause momentarily. When the Measure menu appears, scroll down to the S21 button and click it. See Figure 5-1. Figure 5-1 Specifying an S Measurement 5 -8 Agilent Technologies Z5623A Option H48...
Page 77
“Entry:”. (The white rectangle is referred to as the Entry field.) A text entry cursor (a vertical line) will then appear in the white rectangle. Figure 5-3 Title Entry Window Agilent Technologies Z5623A Option H48 5-2. Figure Performance Verification Setting Up Limit Testing 5-3.
Page 78
Step 5. Click Scale on the menu bar. When the Scale menu appears, scroll down to the Scale... button and click it. See Figure 5-5 The “Scale...” Button 5 -10 Figure 5-4. You can Figure 2-9 on Figure 5-5. Agilent Technologies Z5623A Option H48...
Page 79
5-7. Click the OK button. Figure 5-6 The Scale Window With Default Values Figure 5-7 The Scale Window With Desired Values For Insertion Loss Agilent Technologies Z5623A Option H48 Performance Verification Setting Up Limit Testing Figure 5-6 will pop up. When it does, set...
Step 3. When the Limit Table pops up as shown in from Table 5-4 shown in Figure 5 -12 Figure to the Limit Table on the network analyzer. The result is 5-11. Agilent Technologies Z5623A Option H48 Figure 5-8. Figure 5-9, click the 5-10, transfer entries...
Page 81
Figure 5-10 Empty Limit Table Table 5-4 Limit Table Entries for Insertion Loss Segment Number Figure 5-11 Completed Limit Table for Insertion Loss Agilent Technologies Z5623A Option H48 Limit Beginning Ending Type Stimulus Stimulus 300 KHz 1.3 GHz 1.3 GHz...
Electronic Calibration (ECal) module. Notice that the ECal module does not meet the insertion loss specifications of the test set. 5 -14 Figure 5-9 will pop up. 5-12. 5-13. The trace shown in the figure is an S Agilent Technologies Z5623A Option H48 Figure 5-8.
Step 1. Click File on the menu bar. When the File menu appears, scroll down to the Save As... button and click it. See Figure 5-14 “Save As...” Button Agilent Technologies Z5623A Option H48 Performance Verification Setting Up Limit Testing Figure 5-14.
Page 84
“Save As...” Window Figure 5-16 Specifying The Name Of The Limit Test File For Insertion Loss 5 -16 Figure Table 4-1 page 4-13. For example, if the Table 5-3 on page Agilent Technologies Z5623A Option H48 5-15, type a file Figure 5-7.
Step 3. Recall the calibration file listed in “How to Recall a should look something like Figure 5-17 Insertion Loss Test Setup, Cal File Recalled Agilent Technologies Z5623A Option H48 Table 4-1 Calibration” on page 4-14.) The resulting display Figure 5-17.
Page 86
Step 4. Recall the limit line file for insertion loss. The file name should be listed Table 5-3 on page 5-7. The resulting display should look something like Figure 5-18. Figure 5-18 Insertion Loss Test Setup, Limit Test File Recalled 5 -18 Agilent Technologies Z5623A Option H48...
Title Entry Window With “Return Loss (Port Active)” Entered Step 4. Click Scale on the menu bar. When the Scale menu appears, scroll down to the Scale... button and click it. See Agilent Technologies Z5623A Option H48 Table 4-1 on page Calibration” on page 5-2.
Electronic Calibration (ECal) module. Notice that the ECal module does not meet the return loss specifications of the test set. Figure 5-22 Return Loss (Port Active) Limit Test On ECal Module Agilent Technologies Z5623A Option H48 Figure 5-9 Figure 5-12. Figure 5-22.
Table 4-1 on page 5-16. After entering the file name, record it 5-7. Next, click the Save button. Table 4-1 on page page 4-14.) The resulting display Figure 5-17. Agilent Technologies Z5623A Option H48 5-15, type a file 4-13. For 4-13. (Consult...
Page 91
Step 4. Recall the limit line file for return loss (port active). The file name should be listed in Table 5-3 on page 5-7. The resulting display should look something like Figure 5-23. Figure 5-23 Return Loss (Port Active) Test Setup, Limit Test File Recalled Agilent Technologies Z5623A Option H48 5 -23...
Recalling the Limit Test File Recall the limit test file that you have just saved. (Consult Recall a Calibration” on Recall” list. Agilent Technologies Z5623A Option H48 Figure 5-9 Figure 5-16. After entering the file name, record it in page 4-14.) This will put it on the “Quick...
Step 1. Press the Preset button. Step 2. Click Channel on the menu bar. When the Channel menu appears, scroll down to the Power... button and click it. See Figure 5-25 The “Power...” Button 5 -26 Figure 5-25. Agilent Technologies Z5623A Option H48...
Page 95
Power to +10 dBm as shown in Power window. Figure 5-26 Power Window with Default Values Figure 5-27 Power Window with Values Desired for Crosstalk Agilent Technologies Z5623A Option H48 Performance Verification Setting Up Limit Testing Figure 5-26 will pop up. Set the Test Port Figure 5-27.
IF Bandwidth to 100Hz. (Type, “100<space>”, without the quotation marks.) The result is shown in button. Figure 5-29 IF Bandwidth Window with Default Value 5 -28 Figure 5-28. Figure 5-29, Figure 5-30. Click the OK Agilent Technologies Z5623A Option H48...
Page 97
Step 3. Click Channel on the menu bar. When the Channel menu appears, scroll down to the Average... button and click on it. See Figure 5-31 The “Average...” Button Agilent Technologies Z5623A Option H48 Performance Verification Setting Up Limit Testing Figure 5-31.
Page 98
Average Factor by using the up arrow.) The results are shown in Figure Figure 5-32 Average Window with Default Settings Figure 5-33 Average Window with Settings Desired for Crosstalk 5 -30 Figure 5-32 5-33. Click the OK button. Agilent Technologies Z5623A Option H48 appears, click the...
Page 99
Status Bar button and click it as shown in Figure 5-34. The status bar will appear at the bottom of the screen as shown in Figure 5-35. Figure 5-34 The “Status Bar” Button Agilent Technologies Z5623A Option H48 5 -31...
Step 5. Click Scale on the menu bar. When the Scale menu appears, scroll down to the Scale... button and click it. See 5 -32 5-2. Figure Agilent Technologies Z5623A Option H48 5-3. When this window pops Figure 5-5.
Page 101
• Scale Per Division = 20 dB • Reference Level = 0 dB • Reference Position = 9 Divisions When these values have been entered, the Scale window will look like Figure 5-37. Click the OK button. Agilent Technologies Z5623A Option H48 5 -33...
5-7. Next, click the Save button in the Save As window. Recalling the Limit Test File Recall the limit test file that you have just saved. (Consult Recall a Calibration” on Recall” list. Agilent Technologies Z5623A Option H48 Figure 5-9 will pop up. Figure 5-14. Figure page 4-14.) This will put it on the “Quick...
Record the results in 5 -36 E8358A (300 kHz - 9.0 GHz) Port 1 Port 2 Reflection for a list of commands. Table 5-8. 5-7. Table Agilent Technologies Z5623A Option H48 Figure 5-39. Transmission Table 5-3 on Table 5-8.
Step 7. Move the RF cable to the next test port. Repeat Step 8. Repeat Step 7 until all eight test ports have been measured. Agilent Technologies Z5623A Option H48 Performance Verification Verifying Return Loss and Insertion Loss Specifications Table 5-8.
Page 106
Pass/Fail Test Results For Insertion Loss And Return Loss Test Port 5 -38 Connected To Reflection Port Terminated Insertion Return Loss Return Loss Loss (Port Active) Agilent Technologies Z5623A Option H48 Internally Connected To Transmission Port Insertion Return Loss (Port Off) Loss (Port Active)
Transmission Port to the other test port having a short. Step 8. Read the results of the crosstalk test on the screen of the network analyzer. Record the results in Agilent Technologies Z5623A Option H48 E8358A (300 kHz - 9.0 GHz)
Page 108
1 and 2 2 and 3 3 and 4 4 and 5 5 and 6 6 and 7 7 and 8 5 -40 Step 7 Step until all seven adjacent pairs of test ports have been Result Agilent Technologies Z5623A Option H48...
• Return loss on the DUT's input and 2 output ports (A and B) • Insertion loss (or gain) between the DUT's input and port A • Insertion loss (or gain) between the DUT's input and port B 6 -2 Figure 6-1, the following tests will be Agilent Technologies Z5623A Option H48...
Page 111
Do not use the test set to change the RF signal path direction when you CAUTION are using a full two-port calibration. Doing so will render the calibration invalid. Instead, use the internal transfer switch in the analyzer. Agilent Technologies Z5623A Option H48 (300 kHz - 9.0 GHz) Port 1 Reflection Input...
Page 112
Measuring Multiport Devices Calibrating the Test System 6 -4 Agilent Technologies Z5623A Option H48...
Advanced Topics This chapter contains the following sections: • Using an External Computer to Control the Test Set • Using the Control Lines Connector Agilent Technologies Z5623A Option H48 7 -1...
Number of times Switch #13 has operated Number of times Switch #14 has operated Number of times Switch #15 has operated Number of times Switch #50 has operated Number of times Switch #51 has operated Agilent Technologies Z5623A Option H48 3-3. In...
GPIB Interface and Visual Basic, use the following commands to disable EOI and EOL, send the necessary data to the test set, and re-enable EOI and EOL. Be sure to re-enable EOI and EOL before sending data to another NOTE instrument. Agilent Technologies Z5623A Option H48 7 -3...
If you are using HPVEE, be sure to set the Direct I/O in the Advance Device Configuration so the Read Terminator and Write EOL Sequence is set to “\n”. In the I/O Transaction make sure the EOL is ON. 7 -4 Agilent Technologies Z5623A Option H48...
If this attribute is set to , the bit terminates read operations. VI_FALSE Write Commands: Append all commands with “\n,” for example, *rst\n Read Commands: The Z5623A Option H48 returns data terminated by \r\n. Agilent Technologies Z5623A Option H48 7 -5...
OUTPUT 712;"5;" ! sets C5-C1 to OOLOL Table 3-3 on page 3-10 lists all possible combinations of control line states. For each combination, the table lists the binary and decimal numbers required to command that combination. 7 -6 Agilent Technologies Z5623A Option H48...
(The maximum allowable is 250 mA per control line.) The values computed in this paragraph do not depend on V , the voltage used to control the DUT. (V is applied to pin 3 in Figure 7-2.) Agilent Technologies Z5623A Option H48 7 -7...
Page 120
Switch Driver Circuit For Use With Control Lines Connector 10 kΩ C1_in (from Control Line Connector on Test Set) 7 -8 +15 V (from Power Supply) Temic/Siliconix DG413 -15 V Agilent Technologies Z5623A Option H48 C1_out (to DUT) 10 V...
Page 121
Power Supply can be adjusted from +1.2 to +12.0 V if Power Supply = +15 V Agilent Technologies Z5623A Option H48 ) ≤ 40 V ) + I ) from 1.2 V to 12.0 V provided that (a) a power supply...
Page 122
Advanced Topics Using the Control Lines Connector 7 -10 Agilent Technologies Z5623A Option H48...
Service This chapter contains the following sections: • Adjustments • Theory of Operation • Troubleshooting • List of Replaceable Parts Agilent Technologies Z5623A Option H48 8 -1...
+24 V, +12 V, +5 V. The power LED on the front panel indicates that the instrument is on and that the power supply is providing power. Agilent Technologies Z5623A Option H48 8 -3...
Control signals and DC power are provided by a cable connected to the main switch board. Connector Replacement The 50 Ω Type-N connectors are available separately. It is possible to replace them in the field. 8 -4 Agilent Technologies Z5623A Option H48...
• Finally, disconnect the DC power cable from the power supply to the main switch board and measure the voltages. They should be +15V, +12V, and +5V. If not, replace the power supply. Agilent Technologies Z5623A Option H48 Service Troubleshooting...
Reflection or Transmission path. Check each of the RF paths for connection. If an RF path is not connected to the necessary port or terminated in 50 Ω, replace the controller and switch driver board. 8 -6 Agilent Technologies Z5623A Option H48...
If any of these parts or assemblies is replaced, you must perform all performance tests to verify conformance to specifications. Table 8-1 Replaceable Parts for Z5623A Option H48 Reference Designator Sw50-Sw57 Sw10-Sw15 Agilent Technologies Z5623A Option H48 Description Part Number 24 Volt Power 0950-2252 Supply 2x16 LCD 2090-0370 1P2T Switch 26.5...
Page 130
(CT / PGU) Support Group at (707) 577-6802 with the part number, module/model number, and option number. If ordering parts through your local Agilent Technologies Sales and Service Office, specify that they are ordered through the Component Test PGU Support Group.
Intentional interruption is prohibited. The power cord is connected to internal capacitors that may retain dangerous WARNING electrical charges for 5 seconds after disconnecting the plug from its power supply. Agilent Technologies Z5623A Option H48...
4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used. Agilent Technologies Z5623A Option H48 Safety and Regulatory Information Safety Information...
Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation. product (provided Agilent Technologies Z5623A Option H48...
Deutschland). Laermangabe (Noise Declaration) Geraeuschemission (Acoustic Noise Emission) LpA < 70 dB am Arbeitsplatz normaler Betrieb (at Operator Position during Normal Operation) nach DIN 45635 t.19 (per ISO 7779) Agilent Technologies Z5623A Option H48 Safety and Regulatory Information Regulatory Information...
Page 136
Safety and Regulatory Information Regulatory Information Agilent Technologies Z5623A Option H48...