These servicing instructions are for use by qualified personnel only. To avoid
WARNING
electrical shock, do not perform any servicing unless you are qualified to do
so.
WARNING
The opening of covers or removal of parts is likely to expose dangerous
voltages. Disconnect the instrument from all voltage sources while it is being
opened.
This product is designed for use in Installation Category II and Pollution
WARNING
Degree 2 per IEC 1010 and 664 respectively.

Cautions

Cautions applicable to this instrument are:
Always use the three-prong ac power cord supplied with this
CAUTION
instrument. Failure to ensure adequate earth grounding (by not using
this cord) can cause instrument damage.
This instrument has autoranging line voltage input; be sure the supply
CAUTION
voltage is within the specified range.
Ventilation Requirements: When installing the instrument in a cabinet,
CAUTION
the convection into and out of the instrument must not be restricted.
The ambient temperature (outside the cabinet) must be less than the
maximum operating temperature of the instrument by 4 °C for every
100 watts dissipated in the cabinet. If the total power dissipated in the
cabinet is greater than 800 watts, forced convection must be used.
Agilent Technologies Z5623A Option H48
Safety and Regulatory Information
Safety Information
9-3