AC Current Output

To set an ac current output:
1. Make sure the Calibrator is in standby (STANDBY annunciator lit). Push  if
necessary.
2. If the UUT is not connected, connect it now as described earlier in this chapter under
"Connect the Calibrator to a UUT."
3. Set the UUT to measure ac current on the appropriate range.
4. Enter a current magnitude with the numeric keypad.
5. Push  or  if necessary.
6. Push .
7. The display now shows the amplitude of the entry. If an entry error is made, push 
to clear the display, then reenter the value. If the most recent digit entered is in error,
push to clear that digit. The illustration of the display below assumes an entry of
125 mA: