Calibration Check Shift Results; Raw Data Results - Fluke 5730A/03 Operator's Manual

Multifunction calibrator
Hide thumbs Also See for 5730A/03:
Table of Contents

Advertisement

Calibration Check Shift Results

The Calibration Check report presents changes in the magnitude of self-correction
generated by calibration check. Calibration check uses internal standards rather than
external standards as the basis for comparison, and does not make permanent changes to
calibration constants. This report is available only after a calibration check has been
performed and until the Calibrator is reset. The report contains the subsequent
information, and is shown below:
Heading that shows the date and time the report was saved and user report string
Modules present (installed or attached)
Most recent calibration dates and temperatures
Values of externally-calibrated internal references
DC voltage output shifts for each range
AC voltage output shifts for each range
AC and dc current output shifts for each range
Resistance shifts for each value
5725A Boost Amplifier output shifts (if attached). The calibration report file name is
CKyymmdd.CSV, where yymmdd is the date today.

Raw Data Results

This report gives the values for the internally stored correction factors (calibration
constants), and is useful primarily for our service centers.
Operator Maintenance and Calibration
Save Calibration Reports
7
7-23