Fluke 5730A/03 Operator's Manual page 142

Multifunction calibrator
Hide thumbs Also See for 5730A/03:
Table of Contents

Advertisement

5730A
Operators Manual
ADDR
Description
Parameter
Example
ADDR ?
Description
Parameter
Response
Example
ADJOUT?
Description
Parameter
Response
Example
BOOST
Description
Parameter
Example
6-16
Table 6-5. Commands
Sequential command. Ignored if not in remote. Sets the GPIB interface bus address.
Bus address
ADDR 4
Sets the GPIB interface bus address to 4
Sequential command. Gets the GPIB interface bus address.
None
Integer
ADDR?
Returns 4 if the gpib interface bus address is set to 4
Sequential Command.
Returns the adjusted output magnitude and frequency. The adjusted output magnitude is
the output after modification by the front panel knob or the INCR remote command. In all
output functions but resistance, ADJOUT? behaves exactly like OUT?. In the resistance
function, OUT? returns the actual Calibrator output, which cannot be adjusted, and
ADJOUT? returns what would be the reading on the Display in direct operation. The
frequency is always the present output frequency.
None
(Float) Output amplitude
(String) Units (V, DBM, A, or OHM)
(Float) Frequency (0 if dc or ohms)
1.256983E+01,V,0
(12.56983V)
1.883E-01,A,4.42E+02
(188.3 mA, 442 Hz)
1.9E+06,OHM,0
(1.9 MΩ)
Overlapped Command, ignored if not in remote.
Activates and deactivates an accessory amplifier.
ON
Activates the appropriate amplifier for the type of output)
OFF
Deactivates the active amplifier
BOOST ON
Activates the appropriate amplifier if the last OUT command selected an output
supported by the attached amplifier.