FORMAT
Sequential command, ignored if not in remote.
Use with extreme care. Restores the contents of the non-volatile memory to factory
Description
defaults. The non-volatile memory holds calibration constants and setup parameters. All
calibration data is lost permanently. The Calibrator security state must be set to OFF: or
an execution fault occurs.
ALL
(Replaces the whole contents with defaults)
B5725 (Replaces 5725A calibration constants with defaults)
Parameter
CAL
(Replaces all calibration constants with defaults)
RNG
(Replaces range calibration constants with defaults)
SETUP (Replaces setup parameters with defaults)
GWADDR
Sequential command. Ignored if not in remote. Sets Ethernet gateway address for LAN
Description
communication when NOT in DHCP mode.
Gateway address (quoted string consisting of 4 decimal values bound between 0-255
Parameter
separated by periods).
GWADDR "129.196.136.1"
Example
Sets the Etherent gateway address to 129.196.136.1
GWADDR?
Sequential command. Returns Ethernet gateway address for LAN communication when
Description
NOT in DHCP mode. When in DHCP mode, the response will be default.
Parameter
None
Response
String
GWADDR?
Example
Returns 129.196.136.1 if the gateway address is set to 129.196.136.1 and DHCP is not
enabled. Returns default if DHCP is enabled.
ID52120? (returns the number of 52120s connected and their serial numbers)
Sequential command. Returns the number of 52120s connected and their serial
Description
numbers.
Parameter
None
Response
One or more integers
ID52120?
Example
Returns 0 if no 52120As are connected. Returns 2, 7346432, 8883213 if two 52120A's
are connected with the indicated serial numbers.
Table 6-5. Commands (cont.)
Remote Commands and Syntax
6
Commands
6-31