Fluke 5730A/03 Operator's Manual page 173

Multifunction calibrator
Hide thumbs Also See for 5730A/03:
Table of Contents

Advertisement

LOCAL
Sequential command.
Description
Enables the local state. This command duplicates the IEEE-488-GTL (Go to Local)
message.
Parameter
None
LOCKOUT
Sequential command.
Description
Enables the local lockout state. This command duplicates the IEEE-488 LLO (Local
Lookout) message.
Parameter
None
REMOTE
Sequential command.
Description
Enables the remote state. This command duplicates the IEEE-488 REN (Remote
Enable) message.
Parameter
None
Function
DC220MV
DC Voltage
DC1100V
DC220UA
DC Current
DC5725A
AC2_2MV
AC Voltage
AC220V
AC220UA
AC Current
AC5725A
OHM0
OHM100
Resistance
OHM19K
OHM10M
WB1_1MV
Wideband
WB330MV
Table 6-6. Serial Remote Control Commands
Table 6-7. Range Identifiers for Remote Commands
DC2_2V
DC2_2MA
DC52120A_2A
AC22MV
AC1100V
AC2_2MA
AC52120A_2A
OHM1
OHM190
OHM100K
OHM19M
WB3_3MV
WB1_1V
Remote Commands and Syntax
Ranges
DC11V
DC22V
DC22MA
DC220MA
DC52120A_20A
DC52120A_100A
AC220MV
AC2_2V
AC5725A
AC22MA
AC220MA
AC52120A_20A
AC52120A_120A
OHM1_9
OHM10
OHM1K
OHM1_9K
OHM190K
OHM1M
OHM100M
WB11MV
WB33MV
WB3_5V
6
Commands
DC220V
DC2_2A
AC22V
AC2_2A
OHM19
OHM10K
OHM1_9M
WB110MV
6-47