Fluke 5730A/03 Operator's Manual page 145

Multifunction calibrator
Hide thumbs Also See for 5730A/03:
Table of Contents

Advertisement

CAL_DATE?
Sequential command.
Description
Returns the date of the most recent calibration of the specified type.
B5725
CAL
Parameter
WBFLAT
WBGAIN
ZERO
(Integer) Date as MMDDYY, DDMMYY, or YYMMDD depending on the setting of
Response
DATEFMT.
CAL_DAYS?
Sequential command.
Description
Returns the number of days elapsed since the last calibration activity of the specified
type.
B5725
CAL
Parameter
WBFLAT
WBGAIN
ZERO
Response
(Integer) Number of elapsed days
CAL_INTV
Sequential command, ignored if not in remote.
Description
Sets the calibration interval for main output calibration. This value is saved in nonvolatile
memory and used for calculating the Calibrator output uncertainty.
Parameter
1, 90, 180, or 365- the number of days in the calibration cycle.
CAL_INTV?
Sequential command.
Description
Returns the calibration interval for main output calibration.
Parameter
None
Response
(Integer) 1, 90, 180, or 365- the number of days in the calibration cycle.
CAL_PASSWD
Sequential command. Ignored if not in remote. Sets security password. The Calibrator
Description
secure state must be set to off or an execution fault results.
1.
current security passcode (quoted string consisting of up to 8 decimal digits).
Parameter
2.
new security passcode (quoted string consisting of up to 8 decimal digits).
CAL_PASSWD "5730", "12345"
Example
Sets the security passcode to 12345.
Table 6-5. Commands (cont.)
(Last 5725A Amplifier calibration)
(Last calibration of the Calibrator)
(Last wideband flatness calibration)
(Last wideband gain calibration)
(Last dc V zero calibration)
(Last 5725A Amplifier calibration)
(Last calibration of the Calibrator)
(Last wideband flatness calibration)
(Last wideband gain calibration)
(Last dc V zero calibration)
Remote Commands and Syntax
6
Commands
6-19