9. Enter a frequency with the numeric keypad followed by  or if necessary. The
10. Push . The Calibrator clears the entry from the entry bar toward the bottom
11. Push  to activate the Calibrator output. The UUT will now respond to the
The Calibrator stays in the ac function after a non-zero frequency is
entered. To change back to dc, enter a frequency of 0 Hz or a signed
voltage.
4-18
display now shows the amplitude and frequency of the entry. If an entry error is
made, use  to clear the most recently entered digit, or push  to clear the
display. Reenter the value. The display shown below assumes an entry of 2.8 kHz.
3/14/2013
10:52am
Standby
19V 2.8kHz
External
Sense
OFF
and shows in the larger area above. No voltage is available at the output terminals
until  is pushed.
applied voltage.
In AC voltage output, the selections available at the bottom are:
External Sense
External Guard
Boost
Wideband
Setup Menu
In addition, the Phase Control and Scale functions are available. The output range
is always automatically chosen and cannot be locked in ac volts.
+/- 0.00050mV (1 year) 213 days since calibration
Auto Range 220 mV
0.00000
External
Guard
OFF
Note
mV
Wideband
Setup
Menu
OFF