Fluke 5730A/03 Operator's Manual page 159

Multifunction calibrator
Hide thumbs Also See for 5730A/03:
Table of Contents

Advertisement

ISCE
Sequential command.
Description
Loads a byte into the Instrument Status Change Enable register described under "Check
the Calibrator Status."
Parameter
The decimal equivalent of the binary number to load into the register.
ISCE 56
Example
Enables bits 3 (BOOST), 4 (RCOMP), and 5 (RLOCK) in the Service Request Enable
register.
ISCE?
Sequential command.
Description
Returns the byte from the Instrument Status Change Enable register, described under
"Check the Calibrator Status."
Parameter
None
Response
The decimal equivalent of the register contents byte.
ISCE?
Example
Returns: "4" if bit 3 (BOOST) is enabled (1) and the rest of the bits are disabled (0). (See
"Check the Calibrator Status" for details.)
ISCR?
Sequential command.
Description
Returns and clears the byte from the Instrument Status Change Register, described
under "Check the Calibrator Status."
Parameter
None
Response
The decimal equivalent of the register contents byte.
ISCR?
Example
Returns: "8" if bit 3 (BOOST) is set (1) and the rest of the bits are reset (0). (See "Check
the Calibrator Status" for details.)
ISR?
Sequential command.
Description
Returns and clears the byte from the Instrument Status Register, described under
"Check the Calibrator Status."
Parameter
None
Response
The decimal equivalent of the register contents byte.
ISR?
Example
Returns: "16" if bit 4 (RLOCK) is set (1) and the rest of the bits are reset (0). (See "Check
the Calibrator Status" for details.)
LCOMP_52120
Overlapped command,. Ignored if not in remote. Sets inductance compensation for
Description
attached 52120As.
ON (enables LCOMP for all attached 52120s)
Parameter
OFF (disables LCOMP for all attached 52120s)
Table 6-5. Commands (cont.)
Remote Commands and Syntax
6
Commands
6-33